Cl-6850

Our company's residual chlorine meter is a micro-type water residual chlorine online measuring and controlling instrument. using advanced non-membrane constant voltage sensor. user do no have to replace diaphragm and pharmacy, high sensitivity, rapid signal response, accurate measurement, stable performance and easy maintenance. Using this series of advanced analytical techniques to ensure the long-term stability and reliability of the instrument. There are English menu operation, 485 communication and other functions. It can continuous monitoring and control of residual chlorine content in aqueous solution for drinking water treatment plants, canning plants, drinking water distribution networks, swimming pools, cooling circulating water, water treatment projects, etc.

Product Details

Product Details

Contact Now

Contact Now

Product Details

Main Features

 

  • * Large screen LCD display, English menu (switching from each other).
  • * Simultaneous display of multiple parameters: residual chlorine value, temperature, output current, etc. are displayed at the same time, which is intuitive and easy to read, and has a range over limit prompt.
  • * Capable for factory reset.
  • * Automatic&manual adjustmen temperature compensation function.
  • * Communicate function: RS-485 communication interface (optional, MODBUS protocol partially compatible). 4-20mA current output can be arbitrarily set.
  • * Photoelectric isolation 4-20mA current output.
  • * The hysteresis amount can be arbitrarily set to avoid frequent switching of the switching relays.
  • * Watchdog function: make sure the meter doesn't crash.
  • * Power off protects>10 years.

Technical Indicators

 

  • * Measuring range: residual chlorine: 0-20.00mg/L(ppm), temperature: 0~60°C.
  • * Resolution: 0.01mg/L, 0.1°C
  • * Accuracy: better than ±1% or ±0.01 mg/L, ±0.5°C
  • * Manual/automatic temperature compensation (0~60°C)
  • * Control interface: two sets of ON/OFF relay contacts, which are divided into high-point and low-point alarm signals. The third group of relays: automatic cleaning control function analog output.
  • * Signal isolated output: optocoupler isolation protection 4~20mA signal output.
  • * Relay: The relay hysteresis is arbitrarily set, and the relay load is 3A 220VAC/24VDC.
  • * Working conditions: ambient temperature is 0~60°C, relative humidity ≤90%.
  • * Output load: Load <500Ω (0-10mA), load<750Ω (4-20mA).
  • * Working voltage: 220VAC10%, 50/60Hz.
  • * Size: 96×105×168 mm.
  • * Opening size: 92×92mm.
  • * Weight: 0.9Kg.
  • * Protection level: IP65

Instrument Installation

4.1 Installation of main unit

The instrument should be installed in a clean, dry, well ventilated, vibration-free location with no corrosive gases around it.

1.User need to open a rectangular cutout on the instrument panel or mounting pane.

Cl-6850

2.Insert the meter into the instrument cabinet and tighten the locking bars.

Cl-6850

4.2 Electrodes

The electrode must be installed in the constant current electrode seat hole of the constant current device if the constant flow device continuously passes water, and ensure that the electrode is always immersed in the water. If the constant current device loses water for a long time, please Take out the sensor and cover it with the original rubber sheath. The rubber sheath must ensure the protective liquid. The specific installation is as follows.

A、Taking out the electrode and carefully push the rubber sleeve of the electrode with your fingernail (note: do not twist it by hand to avoid damaging the electrode) until the rubber sleeve is fully pushed out. Note: If you try to install it temporarily, it is best to save the rubber sleeve and the protective liquid in the sleeve, so that after the test installation, the sampling electrode can cover the rubber sleeve and the protective liquid back to the electrode. The protective solution is slightly corrosive, and if it gets on the skin, it will be rinsed off with water.

B、Inserting the electrode into the constant hole of the constant current device (close to the hole of the water outlet). Be careful not to damage the glass head on the top of the sensor. Be sure to use the raw material tape (threaded part) for waterproof sealing before installation. The on-line monitoring of the measured medium should be kept at a constant flow rate and constant, with a minimum flow of 15 cm3/s.

Cl-6850

Instrument Panel And Connection Instructions

The buttons of front side

Cl-6850

1.MENU  2.DOWN  3.UP  4.ENTER  5.ESC

 

Wiring instructions for the rear panel

Cl-6850

 

P1: H2 relay normally open port

1. EL+:(Black wire)

P2: H2 relay common port

2. IN-:(White transparent wire)

HI NO: High point relay normally open port

3.REF: (Yellow wire)

HI COM: High point relay common port

4.Temp

HI NC: High point relay normally closed port

5.Temp

LO NO: Low point relay normally open port

6.empty

LO COM: Low point relay common port

7. RS485 A

LO NC: Low point relay normally closed port

8. RS485 B

Power: 220V (L)

9. 4~20mA+

Power: 220V (N)

10. 4~20mA-

Attention: If you needs RS-232 communication, please select the appropriate RS-485 to RS-232 device. The RS-485 interface part of this instrument is compatible with MODBUS protocol. For details, please consult the manufacturer or distributor.

1.Make sure the wiring is correct before powering on. The wrong wiring may cause damage to the instrument.

2.Power can't be bypassed from high-power equipment, power line is separated from signal line.

3.The standard cable of the electrode sensor is 3 meters long. If necessary, the user can extend the cable by himself, but the maximum is no more than 15 meters. For the extension cable, a silver-plated three-core low-resistance cable with shielded wire is required. Extension cords are recommended to be connected with a highly insulated junction box.

Meter Function Setting

6.1 Menu structure

Cl-6850

6.2 Main interface and Main menu

Cl-6850 Cl-6850

 

The residual chlorine value in the main interface is the main display, and the temperature value and current value are the sub display. Above is the current residual chlorine measurement, 4.12mA is the current output value.

6.3 CL Cal

Due to the constant pressure method, the zero potential and the electrode slope of the new residual chlorine electrode are basically the same. no calibration is required during normal use. As the use process will gradually change, resulting in aging, which requires a certain period of "calibration" to ensure measurement accuracy. The menu is as follows:

Press the up and down keys on the main menu to select the residual chlorine calibration, press ENTER to enter the residual chlorine calibration interface.

Cl-6850

6.4. CL zero Cal

If you need to calibrate the zero point of the residual chlorine electrode, gently blot the electrode before calibration, place the electrode in the residual chlorine-free water (such as distilled water, purified water, etc.), press the ENTER key to enter the instrument, then press the MENU menu selection key to pop up the cursor, press Up and down key modification

Cl-6850 Cl-6850

 

The default value is 0, usually no need to change, wait for stability and press ENTER to save the data. Press the ESC key to return to the previous menu.

6.5. CL slope cal

If you need to calibrate the slope of the residual chlorine electrode. Enter the residual chlorine slope calibration menu. Before calibration, place the residual chlorine electrode in the residual concentration standard chlorine solution, press the ENTER key to enter the instrument, and then press the MENU menu selection key to pop up the cursor. The default value is 0.20mg/L. Press the up and down keys to change it to already Concentration value, such as 0.30mg/L, wait for the residual chlorine value to stabilize and press ENTER to save the data. After the residual chlorine value is stable (±0.01mg/L), it indicates that it has been marked. Press the ESC key to return to the previous menu.

6.6. Compensate set

Pressing the up and down keys in the main menu to select the parameter setting, press ENTER to enter the parameter setting menu, as shown below, the left picture is the first page, and the right picture is the second page. Press the up and down keys to select each setting. The first item is selected in this section.

Cl-6850 Cl-6850

Press ENTER to enter the compensation setup menu. press the MENU menu selection button to pop up the cursor and move the cursor. press the up and down keys to modify. the temperature is divided into manual or automatic mode. if automatic is selected, the manual value is invalid, and vice versa. when equipped with an NTC thermistor, the true value is measured automatically. if there is no NTC thermistor, you can choose manual input.

Cl-6850

 

Press ENTER to save the data, press ESC to return to the previous menu.

6.7. Alarm set

Select the alarm setting in the parameter setting menu and press the ENTER key to enter the alarm setting menu. Press the MENU menu selection button to pop up the cursor and move the cursor. You can press the up and down keys to modify it.

Cl-6850 Cl-6850

 

In order to avoid the relay from beating or controlling the residual chlorine value, this instrument has this function. The specific operation is as follows:

press the up and down to adjust the hysteresis. That is to adjust the relay hysteresis (customer can adjust in this range according to need, the factory default value is 0.5) After adjusting, press ENTER to store, press ESC to return to the previous menu.

Upper limit relay: will be in the actual measured value higher than the high alarm setting value HIGH value

Time action, when the actual measured value falls below (high point H value - hysteresis H value)

Lower limit relay: It will be operated when the actual measured value is lower than the low alarm set value LOW value, and the actual measured value will rise again when it rises above (low point L value + hysteresis L value). Useful to extend the life of the relay or AC contactor. Therefore, the user must set the high, low and hysteresis according to the actual situation.

6.8. 485 comm Set(optional)

Cl-6850 Cl-6850

 

 

Selecting 485 Communication in the parameter setting menu and press ENTER to enter the 485 communication menu. Press the MENU menu selection button to pop up the cursor and move the cursor.

Press the next button to modify it. The communication address (hexadecimal), press ENTER to store the data, press ESC to return to the previous menu. (Note: Please consult the manufacturer or distributor for specific protocol specifications)

6.9. Current output

The residual chlorine value corresponding to the 4-20 mA output is 0-20 mg/L, but the user can arbitrarily set the corresponding value according to his own requirements to meet the industrial control needs. Press the MENU menu selection button to pop up the cursor, move the cursor, press the up and down keys to modify, press the ENTER key to store the data, press the ESC key to return to the previous menu. Output current (mA): I=16×(C-A)/(B-A)+4

Note: I is the output current value, 4mA≤I≤20mA.

C is the current residual chlorine value measured by the meter, 0.00≤C≤20mg/L.

A is the number corresponding to 4 mA in the setting. B is the value corresponding to 20 mA in the setting.

Cl-6850 Cl-6850

 

6.10. Backlight time

Selecting the backlight time in the parameter setting menu and press the ENTER key to enter the backlight setting menu. Press the MENU menu selection button to pop up the cursor and move the cursor. You can press the up and down keys to modify. Press ENTER to store the data, press ESC to return to the previous level.menu. Backlight control allows the meter to save power, protect the display, and extend life

Cl-6850 Cl-6850

6.11. Reset Factory

In the parameter setting menu, select Restore factory value and press ENTER to enter the factory default menu. Press the MENU menu selection button to pop up the cursor, you can press the up and down keys to modify. Press ENTER to store the data, press ESC to return to the previous menu

Cl-6850 Cl-6850

 

Daily maintenance points

The instrument has been calibrated before leaving the factory and can be put into use directly by the user.

The on-line monitoring of the measured medium should be kept at a constant flow rate and constant, with a minimum flow of 15 cm3/s.General instrument failure rate is low

The company provides local after-sales guarantee within one year from the date of sale, but does not include damage caused by improper use. If repair or adjustment is required, please return it, the freight must be paid by yourself, and the package must be confirmed to avoid shipping. Damaged on the way, the company will repair the damage inside the instrument free of charge. (Note: Electrodes and standard solutions are consumables and are not covered by the warranty. The company will guarantee the quality of the electrodes when they are delivered.)

 

Get in Touch

If you are interested in our products, you can choose to leave your information here, and we will be in touch with you shortly.

* Name

* E-mail

Phone

*Message

FAQS

Frequently Asked Questions (FAQ) About Cl-6850

What does the Cl-6850 meter do?

Down Down

Used for online measurement of constant pressure chlorine content.

How to install Cl-6850 chlorine meter?

Down Down

Choose a location to install the sensor wiring.

How do I maintain the Cl-6850 Chlorine Meter?

Down Down

Clean, calibrate and check wiring regularly.

Cl-6850 Why is the reading inaccurate?

Down Down

Sensor failure or calibration error.
CL-6850.webp                  
       
                

Related Products

right
right
WS-9300/9350

WS-9300/9350

WS-9200/9250

WS-9200/9250

WS-9200/9250

WS-9200/9250

JIRS-803C

JIRS-803C

JIRS-803C3

JIRS-803C3

JIRS-803C4

JIRS-803C4

JIRS-803C5

JIRS-803C5

Soil Sensor

Soil Sensor

Related News

right
right
Guide to Choosing the Right RO Controller for Water Purification

Apr.16,2026

Guide to Choosing the Right RO Controller for Water Purification

Maintaining a high-performance reverse osmosis system requires more than just quality membranes; it demands precise regulation. An ro controller serves as the brain of the entire operation, monitoring critical parameters such as conductivity, temperature, and flow rates to ensure the output water meets stringent purity standards. Without an efficient controller, systems are prone to membrane fouling, inefficient energy use, and inconsistent water quality. In this comprehensive guide, we will explore how these devices optimize water treatment processes and how to select the best model for your specific industrial or commercial needs.

Guide to Water Purity Using RO with TDS Controller Systems

Apr.09,2026

Guide to Water Purity Using RO with TDS Controller Systems

Achieving the perfect balance of mineral content and purity in drinking water is a challenge for many households and industrial facilities. A ro with tds controller offers a sophisticated solution by combining the rigorous filtration of Reverse Osmosis (RO) with the precision of a Total Dissolved Solids (TDS) adjustment system. While standard RO systems remove almost everything, including beneficial minerals, a TDS controller allows users to re-mineralize the water to a desired level. This ensures that the water is not only safe and pure but also tastes better and provides essential nutrients.

Understanding and Optimizing Reverse Osmosis Systems with a Controller

Apr.07,2026

Understanding and Optimizing Reverse Osmosis Systems with a Controller

In the world of water purification, reverse osmosis (RO) systems are a cornerstone for producing high-quality water for various applications, from drinking water to industrial processes. However, an RO system is only as effective as its control system. This is where the reverse osmosis controller comes into play. This critical component manages and optimizes the RO process, ensuring efficient operation, water quality, and system longevity.

Understanding the Vital Role of an RO System Controller for Optimal Performance

Apr.04,2026

Understanding the Vital Role of an RO System Controller for Optimal Performance

Reverse Osmosis (RO) systems are critical for purifying water across various industries, from residential applications to large-scale industrial processes. At the heart of a well-functioning RO system lies the RO system controller. This intelligent device monitors and regulates all key parameters of the RO process, ensuring optimal performance, efficiency, and water quality. This article delves into the functionalities, benefits, and selection criteria for choosing the right RO system controller, particularly focusing on those offered by Watequipments. Understanding its role is crucial for maximizing the lifespan and effectiveness of your RO investment.

Smart Water Treatment Solutions Using PH ORP Controller Systems

Mar.24,2026

Smart Water Treatment Solutions Using PH ORP Controller Systems

In modern water treatment and industrial fluid management, maintaining precise chemical balance is essential for safety, efficiency, and regulatory compliance.

Comprehensive Guide to Water Depth Samplers and Their Applications

Mar.04,2026

Comprehensive Guide to Water Depth Samplers and Their Applications

Accurate water quality monitoring is crucial in various fields, from environmental science and aquaculture to industrial wastewater management. A key tool for this is the water depth sampler. This article provides an in-depth look at water depth samplers, their applications, types, and essential considerations for choosing the right one for your needs. We'll explore the benefits of utilizing this equipment to collect reliable data for analysis and informed decision-making. Selecting the right sampler can significantly improve the precision and efficiency of your water monitoring program. A water depth sampler, also known as a depth-integrated sampler, is a device designed to collect water samples from specific depths within a body of water. Unlike simply dipping a container into the water, a depth sampler allows for collection of a representative sample across a defined depth range. This is critical because water properties (temperature, salinity, pollutants) often vary significantly with depth. The data gathered with a water depth sampler provides a more accurate assessment of water quality than surface samples alone.

Comprehensive Guide to Horizontal Water Sampler Applications and Selection

Mar.03,2026

Comprehensive Guide to Horizontal Water Sampler Applications and Selection

Collecting water samples at varying depths is crucial for environmental monitoring, research, and quality control. The horizontal water sampler from Watequipment.com offers a precise and efficient solution for this task. This article provides an in-depth look at horizontal water samplers, their applications, benefits, and key considerations when choosing the right model. We will explore how these samplers contribute to accurate data collection and informed decision-making in various fields. A horizontal water sampler is a device designed to collect water samples from specific depths within a body of water – rivers, lakes, oceans, reservoirs, and even wastewater treatment plants. Unlike vertical samplers which primarily focus on depth, horizontal samplers prioritize collecting samples at a consistent depth while moving along a horizontal plane. This is essential for assessing water quality variations across a waterway or within a specific zone. They are used in applications ranging from environmental monitoring and scientific research to industrial process control and regulatory compliance.

Comprehensive Guide to Selecting the Right Surface Water Sampler for Your Needs

Feb.28,2026

Comprehensive Guide to Selecting the Right Surface Water Sampler for Your Needs

Monitoring the quality of surface water is critical for environmental protection, public health, and various industrial applications. A surface water sampler is an essential tool for collecting representative samples from rivers, lakes, ponds, and other surface water sources. This article provides a detailed overview of surface water samplers, their types, applications, and key considerations for selecting the right equipment. Accurate water quality data relies on proper sample collection, and choosing the correct sampler is the first step toward achieving reliable results. Surface water samplers come in various designs, each suited to specific sampling needs. Common types include grab samplers, depth-integrating samplers, peristaltic pumps, and automatic samplers. Surface water samplers are used across a broad spectrum of applications, including environmental monitoring, wastewater discharge monitoring, drinking water source monitoring, agricultural runoff analysis, and research studies.

If you are interested in our products, you can choose to leave your information here, and we will be in touch with you shortly.


en_USEnglish